Specification
Wavelength Range 190-1100 nm
Wavelength Accuracy ±0.5 nm
Wavelength Repeatability ±0.2 nm
Spectral Bandwidth 1.8 nm or 4 nm optional
Stability ±0.001 A/h at 500 nm, 0 A
Baseline Flatness ±0.002 A
Photometric Range 0-200 %T; -0.3-3.0 A; -9999-9999 C
Photometric Accuracy ±0.002A (0-0.5A); ±0.004A (0.5-1A); ±0.008A (1-2A); ±0.3% (0-100%T)
Photometric Repeatability ±0.002A (0-0.5A); ±0.004A (0.5-1A); ±0.008A (1-2A); ±0.15% (0-100%T)
Stray Light ≤0.05 %T (220nm, 340nm)
Display LCD
Noise ±0.001A
Software PC with on-line software
Overall Dimension 470x390x225 mm
Weight 20 kg
Catalog No. 9153174164

Split Beam UV Visible Spectrophotometer LSUVS-101PC

Details

• Wavelength range

  1. 190nm-1100nm

• Optical System

  1. Split Double beam optical system
  2. Czerny-Turner diffraction monochromator having 1200 lines/mm grating

• Bandwidth

  1. 1.8nm/4nm

• Wide Measurement Range with ultra-low Stray Light at highest Resolution

  1. With its ultra-low stray light (≤0.05 %T (220nm, 340nm)) is achieved at wavelength range of 190 - 1100 nm

• Flexibility with various sample size

  1. A wide range of standard cuvettes can be use based on the sensitivity or sample volume requirements.

• Sophisticated and simplified sample analysis software

  1. The Sample analysis software deliver scanning, fixed wavelength analysis, quantitative analysis, data collection, storage, export, and reporting.
  2. With additional feature of an easy access USB port available in the unit, which enables results to be stored directly to a USB memory stick for easy transfer of data

• USB data interface with online PC software ensures powerful function and data processing capabilities

• Automatic 4-position cell holder and optional 7-position or 16-position cell holder

• Variety of optional accessories suitable for different applications including reflection sample holder, solid sample holder, water bath and auto sampler

Application

Labocon Split double Beam UV Visible Spectrophotometer LSUVS-100 Series is widely used for analytical testing in various fields like biochemistry, organic chemistry, pharmaceutical analysis, food testing, medicine and health, environmental protection, life sciences, etc.

Your Recent History View All
    • Flow Rate: 0.035-2280 ml/min
    • Speed: 1-600 rpm, reversible
    • Speed Precision: 0.1 rpm
    • Speed Control: 0-5 V, 0-10 V, 4-20 mA (Optional)
    • Operating Temp.: 0-40° C
    • Excitation Source: Ultraviolet Light & White light
    • Transmitting Wavelength: 312 nm
    • CCD Resolution: 1360x1024 pixel
    • Pixel Density: 12-bit
    • Noise: 70 dB
    • Capacity: 96 wellx4 plates
    • Std Plate Speed Range: 300-1500 rpm
    • Deep-well Plate Speed Range: 300-1000 rpm
    • Shaking Orbit: 3 mm
    • Movement: Horizontal